Industrial Series: GS8500-FID

GS8500-FID Process VOC Analyzer (FID)

Flame ionization detector for continuous total VOC measurement

Range
0-10,000 ppmC
Detection
0.1 ppmC
Response
<2s T90
Linearity
>5 decades
Product Overview

Overview

Continuous process VOC analyzer using flame ionization detection for total hydrocarbon measurement. Low-range sensitivity and linear response across the stated VOC range.

Key Highlights
  • Linear response over 5 decades
  • <2s response time
  • Automatic hydrogen flow control
  • EPA Method 25A compliant

Full Technical Specifications

Key Performance

Key Performance metrics for GS8500-FID
Metric Value Status
Range 0-10,000ppmC confirmed
Detection 0.1ppmC confirmed
Response <2 s T90 confirmed
Linearity >5 decades confirmed
Utilities H₂ 30 mL/min; zero air 300 mL/min confirmed

Measurement

Measurement specifications for GS8500-FID
SpecificationValue
Measuring PrincipleFlame Ionization Detection (FID)
Measurement Range0–10,000 ppmC (configurable to 50,000)
Detection Limit0.1 ppmC
Linearity>5 decades
Response Time (T90)<2 s

Physical

Physical specifications for GS8500-FID
SpecificationValue
Dimensions (W×H×D)483 × 310 × 500 mm (19″ 7U rack)
Weight25 kg
Housing MaterialPowder-coated steel, SS316L sample path
Mounting19″ rack or floor-standing enclosure

Electrical

Electrical specifications for GS8500-FID
SpecificationValue
Power Supply100–240 VAC, 50/60 Hz, 150 W
Analog Output2 × 4–20 mA (isolated)
Digital OutputModbus RTU/TCP, RS-485
Relay Outputs4 × SPDT
Display7″ color touchscreen

Environmental

Environmental specifications for GS8500-FID
SpecificationValue
Operating Temperature+5 °C to +45 °C
Storage Temperature-20 °C to +60 °C
Humidity0–90 % RH (non-condensing)
Hydrogen SupplyCarrier grade H₂ at 30 mL/min
Zero Air SupplyHydrocarbon-free at 300 mL/min

Certifications

Certifications specifications for GS8500-FID
SpecificationValue
EPAMethod 25A compliant
EMCEN 61326-1
SafetyEN 61010-1
QualityISO 9001:2015

Where This Analyzer Fits

This model is mapped to the following industry applications and process duties.

Environmental Compliance
Petrochemical
Pharmaceutical

Deployment Environments

Common process environments where this model is evaluated.

Fugitive emission monitoring
Catalytic oxidizer control
Solvent recovery
EPA fence-line monitoring

Certification Scope

Standards are listed with scope and status so engineering review can verify the applicable enclosure, area, and safety case.

Certification scope matrix for GS8500-FID
Standard Scope Variant Status Document
EPA Method 25A Total gaseous organic concentration using FID, listed in features, specifications, certification detail, and FAQ Project variant approved On request
CE / UKCA Approvals listed in existing certification detail Project variant approved On request
EN 61010-1 / EN 61326-1 Safety and EMC standards listed in specifications and certification detail Project variant approved On request
ISO 9001:2015 Quality standard listed in specifications and certification detail Project variant approved On request
CE

Request Documentation

Request technical datasheets, user manuals, certificates, and application notes for this model.

Evidence Notes

case-study

FID boundary-network architecture for continuous VOC compliance records

Application note based on the existing refinery fence-line scenario: FID architecture supports continuous ppmC coverage, fast response, and Method 25A reference-method comparison for VOC monitoring programs. Site result depends on sample-line layout, calibration practice, flare-event handling, and plant data integration.

Technical / Engineering Details

Technical & Engineering Details

Secondary engineering detail — expand each topic for the full measurement, envelope, sample-system, calibration, integration, maintenance and application evidence.

01 Measurement Principle and Limits

How the measurement is bounded

Flame ionization detection measures total hydrocarbon response as ppmC by ion current from a hydrogen-fueled flame.

Rejects

  • PID-style ppb screening where compound-specific response is the primary duty
  • GC-MS speciation when individual compound identification is required

Requires

  • Carrier-grade H₂ fuel gas at the stated flow
  • Hydrocarbon-free zero air at the stated flow
  • Method 25A calibration, interference-check, and data-recording workflow where EPA reporting is required
  • Flame-out detection, hydrogen solenoid closure, and field-enclosure review for outdoor duty

Interferents and Limits

  • Non-organic gases such as CO and CO₂ are outside the hydrocarbon FID response basis
  • Oxygen background and flame stability should be reviewed through the Method 25A interference-check workflow
02 Operating Envelope

Use these limits as selection inputs, then confirm sample condition, ambient exposure, and materials before quotation.

Ambient Temperature
5-45 C
Humidity
0-90 %RH
relative humidity, non-condensing
Wetted Materials
SS316L sample path
Outside This Envelope
  • Condensing sample path without sample-line and enclosure review
  • Operation without H₂ fuel gas and hydrocarbon-free zero air utilities
03 Sample System Boundary

Sampling mode, conditioning components, and exclusions define where the analyzer responsibility ends and the sample system begins.

Sampling Mode
extractive
Boundary Components
  • SS316L sample path
  • H₂ fuel gas supply
  • hydrocarbon-free zero air supply
  • optional NEMA 4X field enclosure with climate control
Boundary Exclusions
  • Outdoor duty requires enclosure, climate-control, and hydrogen-supply management review
04 Calibration & Validation

Calibration method, interval, and traceability are shown only when structured data is available.

Zero / Span Method

EPA Method 25A calibration and interference-check workflow; span gas details are site-specific in v1 source data

Interval

Quarterly jet cleaning and annual detector refurbishment stated in v1 maintenance FAQ; calibration interval not fully stated

Span Gas Traceability

Method 25A calibration workflow stated in v1 FAQ

05 I/O & Integration

Signal outputs and communication interfaces shown from the published specification fields on this product page.

Analog Output

2 × 4–20 mA (isolated)

Digital Output

Modbus RTU/TCP, RS-485

Relay Outputs

4 × SPDT

DAHS Review Note

Confirm protocol map, channel naming, alarm states, and reporting format during project integration review.

06 Maintenance & Spares site-specific

Tasks, consumables, and access items are shown only when structured maintenance data is available.

Service Tasks
  • Quarterly jet cleaning
  • Annual detector refurbishment
  • Hydrogen fuel-gas and zero-air utility review
  • Flame-out interlock and hydrogen solenoid review for outdoor or field-enclosure duty
Consumables
  • Hydrogen fuel gas
  • Hydrocarbon-free zero air
  • FID jet service parts
07 Application Evidence

Application context from complex industrial environments.

Case Study Petrochemical

Application Context

Refinery fence-line VOC compliance under EPA Method 25A is repeatedly compromised by PID networks saturating above ~2,000 ppm during flare events, leaving the regulator reference test and plant-operated data inconsistent for extended periods.

Architecture Response

An 8-unit GS8500-FID boundary network delivers continuous 0–10,000 ppmC coverage with sub-2 s response and linear behavior across 5 decades, so flare events resolve on the same sample line as the routine stream and the reference-method test can be compared against a continuous, unsaturated plant-operated FID record.

Reference Architecture 8-unit FID network · 0–10,000 ppmC continuous
Selection Questions

Frequently Asked Questions

Why choose FID over PID for continuous monitoring?

FID provides linear response over 5 decades (0.1–100,000 ppm), carbon-number response across hydrocarbons, and a flame-ionization path that avoids UV-lamp aging drift. PID is better for ppb-level screening but saturates above ~10,000 ppm.

Does FID require hydrogen gas?

Yes, the flame requires carrier-grade hydrogen at ~30 mL/min and hydrocarbon-free zero air at ~300 mL/min. H₂ consumption is approximately one standard cylinder per 3 months.

What is EPA Method 25A?

EPA Method 25A defines the procedure for measuring total gaseous organic concentration using FID. The GS8500-FID meets all Method 25A requirements including calibration, interference check, and data recording.

Can it operate outdoors?

Yes, with the optional NEMA 4X (IP66) field enclosure that includes climate control, hydrogen supply management, and flame-out auto-reignition.

What happens if the flame goes out?

The analyzer detects flame-out within 2 seconds and auto-reignites. If reignition fails after 3 attempts, it triggers a fault alarm and closes the hydrogen solenoid valve.

What is the annual maintenance cost?

Approximately $800–1200/year: quarterly jet cleaning, annual detector refurbishment, hydrogen fuel gas (~30 mL/min continuous) and zero air supply. Hydrogen cylinders are the primary recurring OPEX — there are no wet-chemistry reagents but H₂ is consumed continuously.

Review GS8500-FID against site conditions

Send gas range, sample temperature, pressure, moisture, and certification needs before final model selection.