Industrial Series: GS8210-EC

GS8210-EC Portable SO₂/NOₓ Analyzer

Handheld emission checker for SO₂ and NOₓ spot measurements

SO₂ Range
0-2000 ppm
NOₓ Range
0-1000 ppm
Weight
1.1 kg
Battery
8 hours
Product Overview

Overview

Portable electrochemical analyzer for rapid on-site SO₂ and NOₓ measurement. Ideal for compliance spot checks, maintenance diagnostics, and source testing preparation.

Key Highlights
  • Replaceable sensor cartridges
  • Built-in thermocouple input
  • Automatic temperature compensation
  • SD card data storage

Full Technical Specifications

Key Performance

Key Performance metrics for GS8210-EC
Metric Value Status
SO₂ Range 0-2000ppm confirmed
NOₓ Range 0-1000ppm confirmed
Accuracy ±5% of reading or ±5 ppm confirmed
Response <30 s T90 confirmed
Runtime 8 hours continuous confirmed

Measurement

Measurement specifications for GS8210-EC
SpecificationValue
Measuring PrincipleElectrochemical (dual sensor)
SO₂ Range0–2000 ppm
NOₓ Range0–1000 ppm (NO + NO₂)
Accuracy±5% of reading or ±5 ppm
Response Time (T90)<30 s

Physical

Physical specifications for GS8210-EC
SpecificationValue
Dimensions230 × 110 × 60 mm
Weight1.1 kg (with pump)
HousingABS/rubber overmold IP54
Display3.2″ color LCD

Electrical

Electrical specifications for GS8210-EC
SpecificationValue
BatteryLi-ion 3.7 V, 6000 mAh
Runtime8 hours continuous
ChargingUSB-C
Data StorageSD card (16 GB included)

Environmental

Environmental specifications for GS8210-EC
SpecificationValue
Operating Temperature-10 °C to +50 °C
Storage Temperature-20 °C to +60 °C
Humidity15–90 % RH (non-condensing)
ThermocoupleK-type input for stack temperature

Certifications

Certifications specifications for GS8210-EC
SpecificationValue
IP RatingIP54
EMCEN 61326-1
Drop Test1 m

Where This Analyzer Fits

This model is mapped to the following industry applications and process duties.

Environmental
Power Generation
Industrial Maintenance

Deployment Environments

Common process environments where this model is evaluated.

Emission spot checks
Boiler commissioning
SCR performance testing
Environmental audits

Certification Scope

Standards are listed with scope and status so engineering review can verify the applicable enclosure, area, and safety case.

Certification scope matrix for GS8210-EC
Standard Scope Variant Status Document
CE Portable SO₂/NOₓ analyzer Project variant approved On request
UKCA Approval listed in existing certification detail Project variant approved On request
EN 61010-1 / EN 61326-1 Safety and EMC standards listed in existing certification detail Project variant approved On request
IP54 Housing rating listed in specifications and certification detail Project variant approved On request
CE

Request Documentation

Request technical datasheets, user manuals, certificates, and application notes for this model.

Evidence Notes

case-study

Portable SO₂/NOₓ screening workflow before reference-method testing

Application note based on the existing environmental-consulting scenario: portable electrochemical SO₂/NOₓ measurement supports field screening, commissioning checks, and audit preparation before a formal CEMS or reference-method campaign is selected.

Technical / Engineering Details

Technical & Engineering Details

Secondary engineering detail — expand each topic for the full measurement, envelope, sample-system, calibration, integration, maintenance and application evidence.

01 Measurement Principle and Limits

How the measurement is bounded

Dual electrochemical sensors measure SO₂ and NOₓ for portable spot checks and maintenance diagnostics.

Requires

  • Replaceable SO₂ and NOₓ sensor cartridges
  • Operation within stated ambient temperature and non-condensing humidity limits
  • SD card workflow when CSV records are needed

Interferents and Limits

  • High humidity or condensation outside the stated 15-90% RH non-condensing boundary
  • Operation outside the stated -10°C to +50°C operating temperature range
  • Formal compliance reporting that requires certified reference-method or QAL1 CEMS testing
02 Operating Envelope

Use these limits as selection inputs, then confirm sample condition, ambient exposure, and materials before quotation.

Ambient Temperature
-10-50 C
Humidity
15-90 %
relative humidity, non-condensing
Outside This Envelope
  • Condensing stack sample at the sensor inlet
  • Use as certified reference-method testing equipment
03 Sample System Boundary

Sampling mode, conditioning components, and exclusions define where the analyzer responsibility ends and the sample system begins.

Sampling Mode
portable
Boundary Components
  • portable analyzer with pump
  • K-type thermocouple input
  • SD card data storage
Boundary Exclusions
  • No external extractive conditioning system is declared for this portable SKU
  • Dedicated SO₂/NOₓ configuration only; additional gas sensors are not supported
04 Calibration & Validation

Calibration method, interval, traceability, and audit support are shown only when structured data is available.

Zero / Span Method

Dedicated zero/span method not stated in v1 source data

Sensor Service Life

SO₂ sensor: 18-24 months typical; NOₓ sensor: 24-36 months typical

Span Gas Traceability

Traceability not stated in v1 source data

05 I/O & Integration

Signal outputs and communication interfaces shown from the published specification fields on this product page.

Data Storage

SD card (16 GB included)

Charging

USB-C

DAHS Review Note

Confirm protocol map, channel naming, alarm states, and reporting format during project integration review.

06 Maintenance & Spares site-specific

Tasks, consumables, and access items are shown only when structured maintenance data is available.

Service Tasks
  • Replace SO₂ and NOₓ electrochemical sensor cartridges at the stated service-life interval
  • Use the user-replaceable cartridge workflow stated in v1 FAQ for field maintenance
  • Keep SD card records available for survey handoff when field data export is required
Consumables
  • SO₂ sensor cartridge
  • NOₓ sensor cartridge

Cell Life

SO₂ sensor: 18-24 months typical; NOₓ sensor: 24-36 months typical

07 Application Evidence

Application context from complex industrial environments.

Case Study Environmental Consulting

Application Context

Environmental-consulting firms running SO₂/NOₓ spot checks at 100+ industrial sites annually face mobilization burden when each assessment requires a 2-day CEMS setup before any go/no-go decision.

Architecture Response

GS8210-EC portable analyzers let 2-person field teams pre-screen sites in 30 minutes, so sites passing screening avoid the upstream CEMS mobilization while sites that fail get prioritized for full reference-method testing.

Field Workflow Acceleration 30-minute portable pre-screen vs 2-day CEMS setup
Selection Questions

Frequently Asked Questions

How often should I replace the electrochemical sensors?

SO₂ sensor: 18–24 months typical. NOₓ sensor: 24–36 months. Both are user-replaceable cartridges — no factory service required.

Can I measure stack temperature simultaneously?

Yes, the built-in K-type thermocouple input displays stack temperature alongside SO₂/NOₓ readings, useful for emission calculations at reference conditions.

Is it accurate enough for compliance reporting?

The GS8210-EC is designed for screening and diagnostics, not certified reference method testing. For formal compliance, specify GS-CEMS-200 with QAL1 certification.

How does the SD card data export work?

Remove the SD card and insert into any PC. Data is stored as CSV with timestamp, gas readings, temperature, and GPS tag (if enabled).

Can I add additional gas sensors?

The GS8210-EC is a dedicated SO₂/NOₓ instrument. For multi-gas portable detection (O₂, LEL, CO, H₂S), see GS8610-4G.

Review GS8210-EC against site conditions

Send gas range, sample temperature, pressure, moisture, and certification needs before final model selection.