Industrial Series: ZS-SCS-600

ZS-SCS-600 Cool-Dry Sampling System

Compact cooler-dryer system for clean, dry sample delivery

Dew Point
<5°C (adjustable)
Flow Rate
0.5-3 L/min
Filtration
0.5 µm coalescing
Peristaltic
Auto condensate drain
Product Overview

Overview

Peltier-cooled sample conditioning system that delivers clean, dry gas to analyzers. Removes moisture, particulates, and condensable hydrocarbons while preserving analyte integrity.

Key Highlights
  • Precise Peltier dew point control
  • Coalescing filter removes aerosols
  • Automatic peristaltic condensate drain
  • Compact wall-mount enclosure

Full Technical Specifications

Key Performance

Key Performance metrics for ZS-SCS-600
Metric Value Status
Conditioned dew point <5C confirmed
Dew-point setpoint range 2-10C adjustable confirmed
Sample flow 0.5-3L/min confirmed
Coalescing filtration 0.5 µm confirmed
Condensate handling Automatic peristaltic drain with 250 mL trap confirmed

Measurement

Measurement specifications for ZS-SCS-600
SpecificationValue
System TypePeltier-cooled sample conditioning
Dew Point<5°C (adjustable 2-10°C)
Flow Rate0.5-3 L/min
Filtration0.5 µm coalescing
CondensateAutomatic peristaltic drain

Physical

Physical specifications for ZS-SCS-600
SpecificationValue
Dimensions400×500×250mm wall-mount
Weight18kg
MaterialsSS316L / Borosilicate glass / PTFE
Condensate Trap250 mL capacity

Electrical

Electrical specifications for ZS-SCS-600
SpecificationValue
Power100-240VAC 50/60Hz 180W
ControllerMicroprocessor with LED status
Temperature Control±0.5°C PID regulation
Alarms2×relay (over-temp fault)

Environmental

Environmental specifications for ZS-SCS-600
SpecificationValue
Ambient+5°C to +45°C
Inlet Gas TempUp to 60°C (higher with pre-cooler)
Inlet Pressure-50 to +500 mbar
Gas CompatibilityNon-corrosive process gases

Certifications

Certifications specifications for ZS-SCS-600
SpecificationValue
EMCEN 61326-1
SafetyEN 61010-1
QualityISO 9001:2015

Where This Analyzer Fits

This model is mapped to the following industry applications and process duties.

Petrochemical
Environmental
Biogas

Deployment Environments

Common process environments where this model is evaluated.

Process gas conditioning
Ambient air sampling
Biogas pre-treatment
Natural gas quality

Certification Scope

Standards are listed with scope and status so engineering review can verify the applicable enclosure, area, and safety case.

Certification scope matrix for ZS-SCS-600
Standard Scope Status
EN 61010-1 electrical safety for the cooler-dryer cabinet approved
EN 61326-1 EMC for the cooler-dryer controller approved
ISO 9001:2015 quality-system listing from the current published certification specification approved
CE

Request Documentation

Request technical datasheets, user manuals, certificates, and application notes for this model.

Evidence Notes

case-study

Biogas cooler-dryer conditioning architecture

Raw digester gas brings saturated moisture and trace H2S into the analyzer front end. The architecture uses a controlled Peltier cooler, coalescing filtration, and automatic condensate removal to separate liquid carryover before the analyzer inlet.

application-note

Cool-dry boundary for reactive gases

Cool-dry conditioning is a fit for clean, non-reactive gases; water-soluble or reactive compounds require dry-basis correction or a heated sampling review to avoid analyte loss in condensate.

Technical / Engineering Details

Technical & Engineering Details

Secondary engineering detail — expand each topic for the operating envelope, maintenance and spares, and application evidence.

01 Operating Envelope

Use these limits as selection inputs, then confirm sample condition, ambient exposure, and materials before quotation.

Ambient Temperature
5-45 C
Dew Point
5 C
adjustable 2-10 C with ±0.5 C PID control per the specifications and FAQ
Corrosives
  • Non-corrosive process gases
  • Mildly corrosive gases with application review
Wetted Materials
  • SS316L
  • Borosilicate glass
  • PTFE
Outside This Envelope
  • Highly corrosive HCl, HF, or SO3 duty; use heated sampling with appropriate wetted materials
  • Inlet gas above 60 C without a pre-cooler
  • Water-soluble gases such as SO2, NH3, or HCl without dry-basis correction or heated sampling review
02 Maintenance & Spares confirmed

Tasks, consumables, and access items are shown only when structured maintenance data is available.

Service Tasks
  • Replace or inspect the 0.5 µm coalescing filter based on aerosol loading
  • Check peristaltic condensate drain operation and trap level during service rounds
  • Verify Peltier setpoint and high-temperature alarm relay after installation changes
  • Review gas solubility before applying cool-dry conditioning to SO2, NH3, or HCl streams
Consumables
  • Coalescing filter element
  • Condensate drain service parts
Service Access

Compact wall-mount enclosure with LED status and condensate trap access

03 Application Evidence

Application context from complex industrial environments.

Case Study Biogas

Application Context

Biogas upgrading facilities conditioning raw digester gas saturated at 38 °C with 2,000 ppm H₂S need cooler-dryer architecture before the analyzer, because moisture and trace H₂S together accelerate sample-cell degradation.

Architecture Response

ZS-SCS-600 with <5 °C dew point and a coalescing filter removes condensate carryover by design, supporting extended analyzer sample-cell service life across digester-gas duty cycles.

Digester-Gas Conditioning <5 °C dew point · coalescing filter
Selection Questions

Frequently Asked Questions

What is the difference between cool-dry and heated sampling?

Cool-dry removes moisture by cooling below dew point — ideal for clean, non-reactive gases. Heated sampling maintains temperature above dew point — required for corrosive or reactive gases (HCl, SO₃, NH₃) that would be absorbed by condensate.

Can it handle corrosive gases?

The ZS-SCS-600 is designed for non-corrosive to mildly corrosive gases. For highly corrosive streams (HCl, HF, SO₃), use the ZS-SCS-800 heated system with Hastelloy wetted parts.

What happens to the removed condensate?

Condensate drains automatically via a peristaltic pump to a collection vessel or floor drain. The 250 mL internal trap provides buffer capacity during high-moisture conditions.

How precise is the dew point control?

The PID-controlled Peltier cooler maintains dew point within ±0.5°C of the setpoint. Adjustable from 2°C to 10°C to optimize for different gas compositions.

Does cooling affect gas measurement accuracy?

For most gases (CO, CO₂, O₂, CH₄), cooling has negligible effect. For water-soluble gases (SO₂, NH₃, HCl), use dry-basis correction or heated sampling to avoid analyte loss in condensate.

Review ZS-SCS-600 against site conditions

Send gas range, sample temperature, pressure, moisture, and certification needs before final model selection.